56B2 anesthesia machine

56B2 anesthesia machine

Add Time: 2017-09-12 12:13:52
Recommendation:

Introduction

introduction

56B2 Anesthesia Machine, designed with state-of-the-art technologies, is one of high utility with excellent performance and complete functions. Both outstanding and inexpensive, it has the best cost performance that can satisfy any customer.

n  The Ventilator adopts microcomputer control with high-definition 5.7” LCD display, integrating the functions of respiratory control and displaying the monitoring parameters. It is versatile in function and excellent in performance.

n  The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose one kind of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.

n  Reliable pneumatic transmission system.

n  Economic and proper configuration of technologies, optional as required by the Customer.

n  Modularized design and upgrading. Upgrading is available as required by the Customer.

 

Part of Anesthesia Machine

Main body                 High-strength engineering plastic rack, light,

beautiful and corrosion resistant

Scope of application          adult, child

Gas source                  O2: 0.280.6MPa  N2O: 0.280.6MPa

Flow meter                   O2: 0.0510L/min    N2O: 0.0510L/min

O2, N2O linkage and N2O Stopper

 When using nitrous oxide, oxygen concentration>25%;

 When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off.

Flow rate of rapid oxygen supply  2575L/min

Low oxygen pressure alarm     There will be sound alarm when the oxygen pressure < 0.2MPa

Vaporizer                   it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 05 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.

Low oxygen pressure alarm     There will be sound alarm when the oxygen pressure0.2Mpa

Respiratory circuit             working mode: all-close, semi-close, semi-open

                            APL≤12.5kpa

Respiratory bellows           bellows for adults, bellows for children

tidal volume range:01500 ml

 

Part of Ventilator

Display mode                High-definition 5.7 LCD screen display

Ventilation mode             IPPV,SIPPV,SIMV, MANUAL

Ventilation function           PEEP, SIGH, IRV

Ventilation parameters         Tidal volume          501500ml

Rate                 299bmp

SIMV rate                 220bpm

I:E                  2:11:8

Inspiratory trigger pressure -10 20 cmH2O

PEEP                220 cmH2O

Pressure limitation      1060 cmH2O

SIGH    1.5 times the inspiratory time

Parameters for ventilation monitoring  tidal volume, minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure

Security Alarm System

Airway pressure alarm          upper limit setting range   560 cmH2O

low limit setting range    050 cmH2O

Minute ventilation volume alarm  upper limit setting range   120 L/min

low limit setting range    019 L/min

Tidal volume                         upper limit setting range   1001500ml

low limit setting range     01400ml

Sustained high-pressure alarm      it will give alarm when stress have consistently

been higher than 2.5 cmH2O

(Suffocation alarm, Power alarm, Intubation off alarm, battery voltage alarm)

Power                           AC 220V   50Hz UPS, Storage battery

Mechanical Arms End-tidal CO2 monitoring

Specific interface of waste anesthesia gas

*Items within brackets are optional.

 